Air heat aging test box installation and commissioning attention to what?
First, the matters needing attention
1. Take care not to force on the glass during transportation, and do not hit the box.
2. After the equipment is grounded, it should be placed in a stable manner. If the ground is uneven, it should be corrected.
3, please do not use the instrument in direct sunlight or hot and humid places.
4. The equipment should be far away from sources of electromagnetic interference, and the grounding wire of the equipment should be effectively grounded.
5. When the equipment is in normal operation, placing the contents in the box should not affect the air circulation to ensure that the temperature inside the box is uniform.
6. When the power supply is AC220V50HZ, it must use a 16A or above three-core socket, and there is a reliable grounding wire. Do not use the two-pin socket without authorization.
7. Don't put the power cord close to the back, and don't let the instrument or other items press on the power cord, so as not to damage the power cord.
Second, the product parameter project K-WKL-B type
Temperature range room temperature ~200°C or 300°C
Temperature fluctuation range ≤±0.5°C
Temperature distribution uniformity ≤±1°C%
Sample turntable speed /
Number of air changes 0 to 20 times/h
Wind speed 0.5~1.5m/s
Studio size 450×450×500mm
Fan power 60W
Power supply voltage AC220V50Hz
Rated power 200°C2.2KW300°C4KW
Weight about 180kg
Suzhou Keitel Instrument Co., Ltd. specializes in the production of air heat aging test chambers, natural gas exchange aging test chambers, forced air aging test chambers, ventilation aging test chambers, high and low temperature test chambers, constant temperature and humidity test chambers, ultraviolet aging test chambers, and ozone aging. Test box. Welcome to call.